Reactivate the HP 8757 System Interface
21. Set the analyzer to SYSINTF ON, the analyzer and swept signal
generator preset.
22. Press (RECALL)
Recall the swept signal generator parameters
from storage register 1
23. On the swept signal generator, press
(
FLTNESS ON
/
OFF
)
(amber
LED on). The power produced at the point where the power
meter/sensor was disconnected is now calibrated at the
frequencies and power level specified above.
l-46 Getting Started Advanced