Note
U, V, and W-band power sensors are not available from
Hewlett-Packard. For these frequencies use the Anritsu
Power Meter with the
(40 to 60
the
(50
to 75
or the
(75 to 110
power sensors. Since
the Anritsu model
Power Meter is not capable of internally
storing power sensor cal factors, you must manually correct the
data entry. Refer to example 2 for information on manual entry of
correction data.
Setup Swept Signal Generator Parameters
5. Turn on the swept signal generator and press (PRESET).
The following occurs:
The source module’s frequency span is displayed on the swept
signal generator.
The swept signal generator’s leveling mode is automatically
changed from internal to “module leveling”.
The source module’s maximum specified power is set and
displayed.
6. Press FREQUENCY (START)
The frequency sweep is set from 26.5 to 40
7.
Press
[
POWER LEVEL
]
The source module power is set to
for maximum power to the device under test.
Access User Flatness Correction Menu
8. Press POWER
Select Fltness Menu.
Select Delete Menu Delete All. This step insures that the
flatness array is empty.
10. Press
Leave the delete menu and return to the previous
soft key menu.
11. Select Enter Freq
to enter 26.5
as
the first correction frequency. Enter 31, 32.5, and 40
to
complete the list. Notice that the frequencies are arbitrarily
spaced.
Getting Started Advanced