The HP
Scalar Network Analyzer
The connections between the analyzer and the HP 8360 B-Series
are similar to the connections between the analyzer and the
HP
The HP 8360 B-Series differs from the HP
in one connection only. It is unnecessary to connect the modulator
drive signal from the analyzer to the source. The HP 8360 B-Series
internally produces the 27.8
modulated signal necessary for AC
mode measurements on the analyzer. The connections from the
HP 8360 B-Series to the analyzer are:
n
Z-AXIS
n
SWEEP OUTPUT
STOP SWEEP IN/OUT
n
HP-IB Interface
Configure the general-purpose HP 8360 B-Series to HP-IB address 19
and network analyzer language for operation with the analyzer. For
information on selecting the instrument address and language refer to
earlier paragraphs in this chapter.
The dedicated HP 8510 versions (HP
HP 83631B) of the
HP 8360 B-Series cannot be used with the HP
The HP 83550 Series Millimeter-wave Source Modules
Refer to “Leveling with MM-wave Source Modules” in Chapter 1 for
information and examples.
The HP 89708 Noise Figure Meter
Connections from the HP 8360 B-Series to the HP 8970B noise figure
meter are identical to those used with the HP
Configure
the HP 8360 B-Series to an address corresponding to the source
address of the HP 8970, typically HP-IB address 19, and network
analyzer language.
3-22 Installation