![HP 8360 B Series User Manual Download Page 349](http://html.mh-extra.com/html/hp/8360-b-series/8360-b-series_user-manual_163548349.webp)
Sets and queries the ALC leveling source selection switch.
The
value is
l
l
POWer : ALC :
?
Sets and queries the state switch of the ALC. The positions are :
ON-normal ALC operation
OFF-open loop ALC mode
When on, the power can be programmed in fundamental units as
selected by the UNIT:POWer command.
When off, the power is no longer calibrated in absolute units and is
set in units of
of arbitrary modulator setting.
l
POWer:AMPLifier:STATE
l
POWer:AMPLifier:STATE?
Sets and queries the state of the amplifier contained in
the doubler (for those models with a doubler installed).
Programming a specific value for POWer : AMPLif ier : STATE sets
POWer:AMPLifier:STATE:AUTOto OFF.
l
POWer:AMPLifier:STATE:AUTO
l
POWer:AMPLifier:STATE:AUTO?
Sets and queries the automatic selection of the doubler amplifier
state. Programming a specific value for POWer : AMPLif ier : STATE sets
POWer : AMPLif ier : STATE:AUTO to OFF. *RST value is ON.
l
POWer:ATTenuation
l
POWer:ATTenuation?
Sets and queries the output attenuation level. Note that when
setting the attenuator level to 10
the output power is
decreased by 10
Programming a specified attenuation sets
POWer:ATTenuation:AUTO OFF.
l
POWer:ATTenuation:AUTO
l
POWer:ATTenuation:AUTO?
Sets and queries the state of the RF attenuator coupling switch.
Programming a specified attenuation sets
POWer:ATTenuation:AUTO OFF.
ON insures that the amplitude level of the ALC is kept within
optimal limits.
OFF the attenuator setting is set to the value of POW: ATT and left
there. The *RST value is ON.
l
POWer:CENTer
[lvl
l
POWer:CENTer?
Sets and queries the center power for power sweep. Default units
(and units for query response) are determined by the UNIT:POWer
command.
S-42 Operating and Programming Reference
Summary of Contents for 8360 B Series
Page 363: ...NF N 2 w z P w D I oa 0 i B 6 7 On s5 n 20 a g 9 i T2 ...
Page 406: ...2b Menu Maps Menu Maps 2b 1 ...
Page 407: ...t t m ALC Bandwidth Select Module Sense Low High Auto AUTO Front Rear None ALC MENU ...
Page 412: ...rMENU SELECT 1 Fault Fault Info1 Info2 Fault Info3 Clear Fault S E R V I C E M E N U 6 ...