MARKER
Active markers are normally displayed as intensified dots on a CRT
display. With
Markers selected, active markers are displayed
as amplitude spikes (an abrupt discontinuity in the sweep trace).
The marker amplitude can be varied. The swept signal generator
displays:
AMPLITUDE MARKER SIZE:
XXXX
Where XXXX
represents an amplitude value. Use the rotary knob, the step keys,
or the numerical entry keys with the dB(m) terminator key to
set the desired value. If a small change is required, the left and
right arrow keys can be used to underline the digit to be changed.
Select
Maskers again to return to the normal intensified dot
representation. See “Specifications” for the range of acceptable
amplitude values. An asterisk next to the key label indicates this
feature is active.
SCPI:
MARKer:AOFF
Analyzer:
AK1 function on,
function off.
“Marker Operation” in Chapter 1
“Setting Up A Typical Sweep, Example Program 3” in Chapter 1
Function Group
MOD (MODULATION)
Menu Map
Description
This
(Option 002 only) scales the amplitude modulation
exponentially. Amplitude modulation lets the RF output of the
swept signal generator be continuously and exponentially varied at
a rate determined by the AM input or at a rate set by
for
internal AM. See “Specifications” for the AM characteristics, input
range, and damage level. An asterisk next to the key label indicates
that this feature is active.
A-18 Operating and Programming Reference
Summary of Contents for 8360 B Series
Page 363: ...NF N 2 w z P w D I oa 0 i B 6 7 On s5 n 20 a g 9 i T2 ...
Page 406: ...2b Menu Maps Menu Maps 2b 1 ...
Page 407: ...t t m ALC Bandwidth Select Module Sense Low High Auto AUTO Front Rear None ALC MENU ...
Page 412: ...rMENU SELECT 1 Fault Fault Info1 Info2 Fault Info3 Clear Fault S E R V I C E M E N U 6 ...