Function Group
SYSTEM
Menu Map
Description
This
sets the standard starting configuration of the swept
signal generator when the
key is pressed, as set by the
manufacturer. An asterisk next to the key label indicates that this
feature is active. The following is a description of the configuration.
n
Start sweep at the minimum specified frequency.
Stop sweep at the maximum specified frequency.
n
Power level set at 0
n
Sweep time set to auto.
n
CONT sweep.
n
Sweep mode ramp.
n
ALC leveling point internal.
n
ALC leveling mode normal.
n
Markers set to activate at the center frequency of the sweep.
n
All function values stored in memory registers 1 through 9 remain
in their previous states.
n
The checksum of the calibration data is calculated, and if an error
is detected, the calibration data in protected memory is used. If
the checksum of the protected data is not correct, then default
values are used an error message (EEROM FAILED, LOST CAL)
is displayed.
Programming Codes
SCPI:
SYSTem:PRESet:TYPE
Analyzer:
IP, which is the same as
See Also
Preset Mode User
“Changing the Preset Parameters” in Chapter 1
P-10 Operating and Programming Reference
Summary of Contents for 8360 B Series
Page 363: ...NF N 2 w z P w D I oa 0 i B 6 7 On s5 n 20 a g 9 i T2 ...
Page 406: ...2b Menu Maps Menu Maps 2b 1 ...
Page 407: ...t t m ALC Bandwidth Select Module Sense Low High Auto AUTO Front Rear None ALC MENU ...
Page 412: ...rMENU SELECT 1 Fault Fault Info1 Info2 Fault Info3 Clear Fault S E R V I C E M E N U 6 ...