COMMAND SUMMARY
l
This causes the swept signal generator to wait for the pending
commands to be completed before executing any other commands.
For example, sending the command:
TSW;*WAI
allows for synchronous sweep operation. It causes the swept signal
generator to start a sweep and wait until the sweep is completed
before executing the next command.
Table S-l. HP 8360
COMMAND SUMMARY
Command
Parameters
A M
AM depth %
AM frequency
waveform
AM depth
AM source
state
AM type
:AM
auto calibrate
auto RF peak
type of det cal
power correction
value
Parameter
Allowed Values
numeric
to 40
extended numeric
[freq suffix]
or
discrete
discrete
discrete
Boolean
discrete
Boolean
Boolean
discrete
extended numeric
[lvl suffix]
Parameter types are explained in the “Getting Started Programming” chapter.
S-18 Operating and Programming Reference
Summary of Contents for 8360 B Series
Page 363: ...NF N 2 w z P w D I oa 0 i B 6 7 On s5 n 20 a g 9 i T2 ...
Page 406: ...2b Menu Maps Menu Maps 2b 1 ...
Page 407: ...t t m ALC Bandwidth Select Module Sense Low High Auto AUTO Front Rear None ALC MENU ...
Page 412: ...rMENU SELECT 1 Fault Fault Info1 Info2 Fault Info3 Clear Fault S E R V I C E M E N U 6 ...