Rack Flange Kit for
Option 908 swept signal generators are supplied with rack flanges and
Swept Signal
the necessary hardware to install them on the swept signal generator
Generators with
after removing the instrument handles. The following table itemizes
Handles Removed
the parts in this kit.
(Option 908)
Table 3-5.
Rack Flange Kit for Swept Signal Generators with
Handles Removed Contents
Quantity
Description
I
2
Rack Mount Flanges
8
Screws
CAUTION
Ventilation Requirements:
When installing the instrument in a
cabinet, the convection into and out of the instrument must not be
restricted. The ambient temperature (outside the cabinet) must be
less than the maximum operating temperature of the instrument
by 4
for every 100 watts dissipated in the cabinet. If the total
power dissipated in the cabinet is greater than 800 watts, then forced
convection must be used.
Installation 3-13
Summary of Contents for 8360 B Series
Page 363: ...NF N 2 w z P w D I oa 0 i B 6 7 On s5 n 20 a g 9 i T2 ...
Page 406: ...2b Menu Maps Menu Maps 2b 1 ...
Page 407: ...t t m ALC Bandwidth Select Module Sense Low High Auto AUTO Front Rear None ALC MENU ...
Page 412: ...rMENU SELECT 1 Fault Fault Info1 Info2 Fault Info3 Clear Fault S E R V I C E M E N U 6 ...