Pulse Width
Function Group
MODULATION
Menu Map
Description
This
lets you set the swept signal generator to apply a slow
rise pulse filter to both internal and external pulse waveforms. This
results in pulses having approximately 2
rise/fall times. An
asterisk next to the key label indicates that this function is active.
Programming Codes
SCPI:
PULM:SLEW
Analyzer:
NONE
See Also
Pulse Menu
Function Group
Menu Map
Description
Programming Codes
See Also
4
This
lets you set a value for the internal pulse generator’s
pulse width. The range of acceptable values is from 1
to 65.5 ms.
The factory preset value is 1 ms. When this feature is active, its
current value is displayed in the active entry area.
SCPI:
PULSe:INTernal:WIDTh <num>[time suffix] or
Analyzer:
NONE
Pulse Menu
Operating and Programming Reference P-23