![HP 8360 B Series User Manual Download Page 165](http://html.mh-extra.com/html/hp/8360-b-series/8360-b-series_user-manual_163548165.webp)
Function Group
MOD (MODULATION)
Menu Map
Description
This
activates the linearly scaled amplitude modulation
function. The amplitude of the RF output changes linearly as a
function of AM input changes. See “Specifications” for the AM
characteristics, input range, and damage level. An asterisk next to
the key label indicates that this feature is active.
Programming Codes
AM:TYPE
AM[:STATE]
Analyzer:
AM1 function on,
function off
See Also
CONNECTORS, (MOD)
“Optimizing Swept Signal Generator Performance” in Chapter 1
Function Group (MOD)
Menu Map
Description
This
(Option 002 only) activates the amplitude modulation
mode for an external source. The AM source is connected to the AM
modulation connector.
When external AM is in effect, the RF output is amplitude
modulated with a rate and depth set by the external source.
Amplitude scaling is controlled by the following softkeys:
AM Type
AM Type
. An asterisk next to the key
label indicates that external AM is active and
is displayed on the
message line.
A-16 Operating and Programming Reference
Summary of Contents for 8360 B Series
Page 363: ...NF N 2 w z P w D I oa 0 i B 6 7 On s5 n 20 a g 9 i T2 ...
Page 406: ...2b Menu Maps Menu Maps 2b 1 ...
Page 407: ...t t m ALC Bandwidth Select Module Sense Low High Auto AUTO Front Rear None ALC MENU ...
Page 412: ...rMENU SELECT 1 Fault Fault Info1 Info2 Fault Info3 Clear Fault S E R V I C E M E N U 6 ...