How to View or Change an HP-IB Address from the Front Panel
Note
To set an HP-IB address from the front panel, the instrument
address on the rear panel HP-IB switch (Figure 3-2)
must
be set to
31 (all
1. Press SYSTEM
Select HP-IB Menu Adrs
Menu.
3. The swept signal generator displays the three address softkeys:
8360 Adrs , Meter Adrs , and Printer Adrs .
4. Select the desired
5. The swept signal generator displays the address selected for that
instrument.
6. If you want to change the address, use the keypad to enter the
desired address (0 to
then press (ENTER).
Remember
If the swept signal generator displays
Rear panel HP-IB address
must
be 31 (11111) in order to change current address
the address on the rear panel HP-IB switch (Figure 3-2) is set to
something other than 31 (all
How to Prevent a Front Panel Change to an HP-IB Address
To disable the address softkeys, set the instrument address on the
rear panel
HP-IB switch (Figure 3-2) to any address
other than 31
(all
How to Set the HP-IB Address on a Swept Signal Generator without a
Front Panel
If your swept signal generator does not have a front panel, set the
address on the rear panel HP-IB switch (Figure 3-2) to the address
you want (factory default is 19).
Mating Connectors
All of the externally mounted connectors on the instrument are
discussed in “CONNECTORS” in Chapter 2. If you are interested in
the HP part number for a connector, see “Replaceable Parts” in the
HP 8360 B-Series Swept Signal Generator/HP 8360 L-Series Swept
CW Generator Service Guide.
10 MHz Frequency
To keep the internal
frequency reference oven at operating
Reference Selection
temperature, the swept signal generator must be connected to ac line
and Warmup Time
power. The swept signal generator requires approximately 30 minutes
to warm up from a
cold
start
before the OVEN display message
goes off. With a stable outside temperature, internal temperature
equilibrium is reached after approximately two hours. For additional
information on warmup times, see Chapter
“Specifications”.
3-6 Installation