Step Dwell
is active it calibrates the frequency at the end of every frequency
band.
Programming Codes
SWEep:CONTrol:STATe
SWEep:CONTrol:TYPE
Analyzer:
NONE
See Also
Step Control Master, Step Sap Menu
Function Group
Menu Map
Description
Programming Codes
See Also
FREQUENCY
2
This
lets you set dwell times for points in the stepped
frequency mode of sweep operation. The dwell time for points in step
frequency sweep may range from 100
to 3.2 s. The actual time
between points is the sum of dwell and phase lock times.
Select
Step Dwell
, then use the entry area to enter the desired
value.
SCPI:
SWEep[:FREQ uency]:DWELl <num>[time suffix] or
Analyzer:
NONE
Step
Menu, Sweep Mode Step
Operating and Programming Reference S-67