![HP 54645A User'S And Service Manual Download Page 234](http://html.mh-extra.com/html/hp/54645a/54645a_users-and-service-manual_166177234.webp)
9
Is the fan running?
If yes, go to "To run the internal self-tests." If not, do the steps below.
The Low Voltage Power Supply has a thermal cut-out circuit. If the fan is
defective, the Low Voltage Power Supply shuts down when it gets too hot for
safe operation.
a
Disconnect the fan cable from the power supply.
b
Measure the fan voltage at the connector on the power supply.
See the following figure for location of the fan connector. If the fan
voltage is approxi8 Vdc, replace the fan. If not, replace the
power supply.
Fan Connector
+5.25 V
adjust
-5.25 V
adjust
fan connector
Figure 7-10
Testing, Adjusting, and Troubleshooting
To check out the oscilloscope
7-45