![HP 54645A User'S And Service Manual Download Page 201](http://html.mh-extra.com/html/hp/54645a/54645a_users-and-service-manual_166177201.webp)
Testing Threshold Accuracy
Instructions for HP 54645D
These instructions apply only to the HP 54645D Mixed-Signal Oscilloscope.
Testing the threshold accuracy verifies the data channel threshold
accuracy specification. You should perform this test every 24 months
or 4000 hours of operation, whichever comes first.
These instructions include detailed steps for testing the threshold
settings of digital channels D0-D7. After testing digital channels
D0-D7, connect and test digital channels D8-D15. Use the detailed
steps for digital channels D0-D7, substituting digital channels D8-D15
for digital channels D0-D7 in the instructions.
Each threshold test tells you to record the voltage reading in the
Performance Test Record located at the end of this chapter. To check
if each test passed, verify that the voltage reading you record is within
the limits listed on the Performance Test Record.
Table 7-2
Equipment Required
Equipment
Critical Specifications
Recommended Model/Part
Digital Multimeter
0.1 mV resolution, 0.005%
accuracy
HP 34401A
Function Generator
DC offset voltage
±
6.3 V
HP 3325B Option 002
BNC-Banana Cable
HP 11001-60001
BNC Tee
HP 1250-0781
BNC Cable
HP 10503A
BNC Test Connector, 8-by-2
7-12
Summary of Contents for 54645A
Page 5: ...iv ...
Page 12: ...1 Getting Started ...
Page 25: ...1 14 ...
Page 26: ...2 Front Panel Overview ...
Page 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Page 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Page 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Page 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Page 189: ...6 14 ...
Page 190: ...7 Testing Adjusting and Troubleshooting ...
Page 242: ...8 Replacement Parts ...
Page 259: ...8 18 ...
Page 260: ...9 Performance Characteristics ...
Page 271: ...9 12 ...
Page 272: ...10 Messages ...