To perform the self-calibration
In this procedure, you load the default calibration factors to give a known
starting point for the firmware calibration.
However, once the default
calibration factors are loaded, you must perform the remainder of the
firmware calibration to maintain the accuracy of the oscilloscope.
Table 7-13
Equipment Required
Equipment
for HP 54645A
Critical specifications
Recommended
Model/Part
Pulse generator
100 kHz, 1 V
p-p
, rise time 10 ns
HP 8112A
Cable
BNC, 3 feet
HP 10503A
Cable
BNC, 9 inches, Qty 2
HP 10502A
Adapter
BNC tee (m) (f) (f)
HP 1250-0781
Adapter
BNC (f-f)
HP 1250-0080
Feedthrough
50
Ω
, BNC (m) and (f)
HP 11048C
Equipment
for HP 54645D
Critical specifications
Recommended
Model/Part
Cable
BNC, 3 feet
HP 10503A
Probe, 10x1
100 MHz
HP 10074A, Qty 2
Cable
16-channel
HP 54620-61801
Banana Ground
Plug
HP 1252-7310
Testing, Adjusting, and Troubleshooting
To perform the self-calibration
7-33
Summary of Contents for 54645A
Page 5: ...iv ...
Page 12: ...1 Getting Started ...
Page 25: ...1 14 ...
Page 26: ...2 Front Panel Overview ...
Page 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Page 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Page 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Page 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Page 189: ...6 14 ...
Page 190: ...7 Testing Adjusting and Troubleshooting ...
Page 242: ...8 Replacement Parts ...
Page 259: ...8 18 ...
Page 260: ...9 Performance Characteristics ...
Page 271: ...9 12 ...
Page 272: ...10 Messages ...