![HP 54645A User'S And Service Manual Download Page 119](http://html.mh-extra.com/html/hp/54645a/54645a_users-and-service-manual_166177119.webp)
To use Autostore
With Autostore or infinite persistence, the scope updates the display with
new acquisitions, but does not erase the results of previous acquisitions.
Rather than being erased by subsequent acquisitions, each pixel memory
location turned on by a previous acquisition is turned into a pixel in
half-bright. Thus, the results of all previous acquisitions are displayed in
half-bright, with the latest acquisition in full-bright.
Autostore is useful in a number of applications, including these:
•
Display the worst-case extremes of varying waveforms.
•
Capture and store waveforms.
•
Measure noise and jitter.
•
Capture events that occur infrequently.
The “Store” indicator is displayed at the right end of the status line when
Autostore is selected.
To Autostore Multiple Events
•
To turn on Autostore, press the
Autostore
key.
•
To turn off Autostore while in run mode, press
Autostore
again.
Run/Stop/Single/Autostore/Erase Operation
To use Autostore
4-36
Summary of Contents for 54645A
Page 5: ...iv ...
Page 12: ...1 Getting Started ...
Page 25: ...1 14 ...
Page 26: ...2 Front Panel Overview ...
Page 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Page 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Page 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Page 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Page 189: ...6 14 ...
Page 190: ...7 Testing Adjusting and Troubleshooting ...
Page 242: ...8 Replacement Parts ...
Page 259: ...8 18 ...
Page 260: ...9 Performance Characteristics ...
Page 271: ...9 12 ...
Page 272: ...10 Messages ...