4
Press
Single
.
Pressing
Single
arms the trigger circuit. When the trigger conditions
are met, data appears on the display, representing the data points that
the oscilloscope obtained with one acquisition. Pressing
the
Single
again re-arms the trigger circuit and erases the display.
Operating Hints
When
Trigger Mode
is set to
Auto
or
Auto Lvl
(auto-single mode), each
time
is pressed the oscilloscope will wait for a valid trigger. If no
trigger is found after about 40 ms, the oscilloscope will force a trigger, acquiring
whatever signal is present. This is a handy mode to look for totally unknown
situations.
When
Trigger mode
set to
Normal
, pressing
will arm the trigger
circuit and do a single acquisition when a valid trigger is seen.
You can use trace/pixel memories to compare several single-shot events. See
the "Saving and Recalling Configuration" section later in this chapter.
Single
Single
Run/Stop/Single/Autostore/Erase Operation
To capture a single event
4-35
Summary of Contents for 54645A
Page 5: ...iv ...
Page 12: ...1 Getting Started ...
Page 25: ...1 14 ...
Page 26: ...2 Front Panel Overview ...
Page 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Page 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Page 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Page 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Page 189: ...6 14 ...
Page 190: ...7 Testing Adjusting and Troubleshooting ...
Page 242: ...8 Replacement Parts ...
Page 259: ...8 18 ...
Page 260: ...9 Performance Characteristics ...
Page 271: ...9 12 ...
Page 272: ...10 Messages ...