Internal Trig
Sensitivity
1
Press
Setup
, then press
Default Setup.
2
Connect the signal generator to channel 1 using a 50-
Ω
feedthrough
at the oscilloscope input.
Channel 1 is labeled "1" on HP 54645A and "A1" on HP 54645D.
3
Verify the trigger sensitivity at 25 MHz and 0.35 divisions.
a
Set the signal generator to 25 MHz and about 100 mV.
b
Press
Autoscale
.
c
Set the time base to 50 ns/div.
d
Set channel 1 to 100 mV/div.
e
Decrease the output of the signal generator until there is 0.35 vertical
divisions of the signal displayed.
The trigger should be stable. If the triggering is not stable, try
adjusting the trigger level. If adjusting the trigger level makes the
triggering stable, the test still passes. If adjusting the trigger does not
help, see "Troubleshooting the Oscilloscope."
f
Record the result on the Performance Test Record as Pass or Fail.
4
Verify the trigger sensitivity at 100 MHz and 1 division.
a
Change the output of the signal generator to 100 MHz and set
amplitude to about 100 mV.
b
Set the time base to 10 ns/div.
c
Decrease the output of the signal generator until there is 1 vertical
division of the signal displayed.
The trigger should be stable. If the triggering is not stable, try
adjusting the trigger level. If adjusting the trigger level makes the
triggering stable, the test still passes. If adjusting the trigger does not
help, see "Troubleshooting the Oscilloscope."
d
Record the result on the Performance Test Record as Pass or Fail.
5
Repeat steps 1 through 4, substituting channel 2 for channel 1 in the
procedure.
Channel 2 is labeled "2" on HP 54645A and "A2" on HP 54645D.
Testing, Adjusting, and Troubleshooting
To verify trigger sensitivity
7-25
Summary of Contents for 54645A
Page 5: ...iv ...
Page 12: ...1 Getting Started ...
Page 25: ...1 14 ...
Page 26: ...2 Front Panel Overview ...
Page 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Page 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Page 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Page 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Page 189: ...6 14 ...
Page 190: ...7 Testing Adjusting and Troubleshooting ...
Page 242: ...8 Replacement Parts ...
Page 259: ...8 18 ...
Page 260: ...9 Performance Characteristics ...
Page 271: ...9 12 ...
Page 272: ...10 Messages ...