To set up and connect the equipment
Instructions for HP 54645D
These instructions apply only to the HP 54645D Mixed-Signal Oscilloscope.
1
Turn on the test equipment and the oscilloscope. Let them warm up
for 30 minutes before beginning the test.
2
Set up the function generator.
a
Set up the function generator to provide a DC offset voltage at the
Main Signal output.
b
Disable any AC voltage to the function generator output, and enable
the high-voltage output.
c
Monitor the function generator DC output voltage with the
multimeter.
3
Using the 8-by-2 test connector and BNC cable assembly, connect
digital channels D0-D7 to one side of the BNC Tee. Connect the
D0-D7 ground lead to the ground side of the 8-by-2 connector.
4
Using a BNC-banana cable, connect the voltmeter to the other side of
the BNC Tee.
5
Connect the BNC Tee to the Main Signal output of the function
generator.
Threshold Test Setup
Measure time
Save/Recall
Entry
LOGIC ANALYZER
54620A
16 CHANNEL 500 MSa/s
STORAGE
TRIGGER
HORIZONTAL
CHANNEL
INPUTS
L i n e
Time/Div
Select
Trigger out
Delay
Position
Ext trigger in
Figure 7-3
Testing, Adjusting, and Troubleshooting
To set up and connect the equipment
7-13
Summary of Contents for 54645A
Page 5: ...iv ...
Page 12: ...1 Getting Started ...
Page 25: ...1 14 ...
Page 26: ...2 Front Panel Overview ...
Page 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Page 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Page 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Page 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Page 189: ...6 14 ...
Page 190: ...7 Testing Adjusting and Troubleshooting ...
Page 242: ...8 Replacement Parts ...
Page 259: ...8 18 ...
Page 260: ...9 Performance Characteristics ...
Page 271: ...9 12 ...
Page 272: ...10 Messages ...