To verify trigger sensitivity
In this test, you verify the trigger sensitivity by applying 25 MHz to the
oscilloscope. The amplitude of the signal is decreased to the specified levels,
then you check to see if the oscilloscope is still triggered. You then repeat
the process at the upper bandwidth limit.
Test limits:
Internal
trigger
1 mV/div to < 10 mV/div
dc to 25 MHz:
1.0 div or 2.0 mV
p-p
25 MHz to 100 MHz:
1.5 div or 3.0 mV
p-p
10 mV/div to 5 V/div
dc to 25 MHz:
0.35 div or 3.5 mV
p-p
25 MHz to 100 MHz:
1.0 div or 10.0 mV
p-p
External
trigger
dc to 25 MHz:
50 mV
p-p
25 MHz to 100 MHz:
100 mV
p-p
Table 7-9
Equipment Required
Equipment
Critical specifications
Recommended
Model/Part
Signal generator
25-MHz and 100-MHz sine waves
HP 8656B Option 001
Power splitter
Outputs differ < 0.15 dB
HP 11667B
Cable
BNC, Qty 3
HP 10503A
Adapter
N (m) to BNC (f), Qty 3
HP 1250-0780
Feedthrough
50
Ω
, BNC connectors (m) and (f)
HP 11048C
Testing, Adjusting, and Troubleshooting
To verify trigger sensitivity
7-24
Summary of Contents for 54645A
Page 5: ...iv ...
Page 12: ...1 Getting Started ...
Page 25: ...1 14 ...
Page 26: ...2 Front Panel Overview ...
Page 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Page 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Page 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Page 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Page 189: ...6 14 ...
Page 190: ...7 Testing Adjusting and Troubleshooting ...
Page 242: ...8 Replacement Parts ...
Page 259: ...8 18 ...
Page 260: ...9 Performance Characteristics ...
Page 271: ...9 12 ...
Page 272: ...10 Messages ...