To adjust the display
The display adjustments are optional and normally do not require
adjustment. You should use this procedure only for the few cases when the
display is obviously out of adjustment.
Table 7-16
Equipment Required
Equipment
Critical specifications
Recommended
Model/Part
Digital multimeter
Accuracy
±
0.05%, 1 mV resolution
HP 34401A
1
Connect the digital multimeter to the end of R901, which is closest to
the fuse.
2
Adjust
+
B for
+
14.00 V.
3
Press
Print/Utility
. Press
Service Menu
, then
Self Tst Menu
and
Display
.
4
Adjust V.HO (vertical hold) for vertical synchronization.
5
Set the intensity control (on the front panel) to mid-range.
6
Adjust Sub Bri (sub bright) to the lowest setting so that the
half-bright blocks on the display are visible.
7
Increase the intensity control to a comfortable viewing level.
This is usually about 3/4 of its maximum range.
Testing, Adjusting, and Troubleshooting
To adjust the display
7-38
Summary of Contents for 54645A
Page 5: ...iv ...
Page 12: ...1 Getting Started ...
Page 25: ...1 14 ...
Page 26: ...2 Front Panel Overview ...
Page 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Page 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Page 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Page 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Page 189: ...6 14 ...
Page 190: ...7 Testing Adjusting and Troubleshooting ...
Page 242: ...8 Replacement Parts ...
Page 259: ...8 18 ...
Page 260: ...9 Performance Characteristics ...
Page 271: ...9 12 ...
Page 272: ...10 Messages ...