To adjust the power supply
The power supply has a
+
5.1 V adjustment and a –5.25 V adjustment. The
other voltages are based on the
+
5.1 V adjustment. In this procedure you use
a multimeter to measure the
+
5.1 V and –5.25 V, and if necessary, adjust the
power supplies to within tolerance.
Table 7-11
Equipment Required
Equipment
Critical specifications
Recommended
Model/Part
Digital multimeter
0.1 mV resolution, accuracy
±
0.05%
HP 34401A
1
Set up the oscilloscope for the voltage adjustment.
a
Turn off the oscilloscope and disconnect power cable.
b
Remove the cover from the oscilloscope. See chapter 8,
“Replacement Parts."
c
Place the oscilloscope on its side.
d
Connect the negative lead of the digital multimeter to a ground point
on the oscilloscope chassis.
e
Reconnect power cable.
f
Turn on the oscilloscope.
Testing, Adjusting, and Troubleshooting
To adjust the power supply
7-30
Summary of Contents for 54645A
Page 5: ...iv ...
Page 12: ...1 Getting Started ...
Page 25: ...1 14 ...
Page 26: ...2 Front Panel Overview ...
Page 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Page 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Page 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Page 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Page 189: ...6 14 ...
Page 190: ...7 Testing Adjusting and Troubleshooting ...
Page 242: ...8 Replacement Parts ...
Page 259: ...8 18 ...
Page 260: ...9 Performance Characteristics ...
Page 271: ...9 12 ...
Page 272: ...10 Messages ...