To use delayed sweep
You can use the delayed sweep window to locate and horizontally expand
part of the main sweep for a more detailed (high-resolution) analysis of
signals. The following steps show you how to use delayed sweep. These
steps are similar to operating the delayed sweep in analog oscilloscopes.
1
Connect a signal to the oscilloscope and obtain a stable display.
2
Press
Main/Delayed
.
3
Press the
Delayed
softkey.
HP 54645D Showing Delayed Sweep
Delayed-sweep
indicator, time
reference
center
Time/Div
for delayed
sweep
Time/Div for
main sweep
These markers
define the beginning
and end of the
delayed sweep
window
Select main or
delayed sweep
Set trigger position
in acquisition
memory
Main sweep
window
Delayed sweep
window
Figure 5-1
Making Measurements with HP 54645A/D Oscilloscopes
To use delayed sweep
5-4
Summary of Contents for 54645A
Page 5: ...iv ...
Page 12: ...1 Getting Started ...
Page 25: ...1 14 ...
Page 26: ...2 Front Panel Overview ...
Page 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Page 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Page 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Page 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Page 189: ...6 14 ...
Page 190: ...7 Testing Adjusting and Troubleshooting ...
Page 242: ...8 Replacement Parts ...
Page 259: ...8 18 ...
Page 260: ...9 Performance Characteristics ...
Page 271: ...9 12 ...
Page 272: ...10 Messages ...