![HP 54645A User'S And Service Manual Download Page 215](http://html.mh-extra.com/html/hp/54645a/54645a_users-and-service-manual_166177215.webp)
External Trig
Sensitivity
6
Verify the external trigger sensitivity at 100 MHz, 100 mV
p-p
and at
25 MHz, 50 mV
p-p
.
Instructions for HP 54645A Only
This procedure applies only to the HP 54645A Oscilloscope.
a
Using the power splitter, connect the signal generator to the
channel 1 input and to the external trigger input. Use the 50-
Ω
feedthroughs at the oscilloscope inputs.
b
Change the output of the signal generator to 100 MHz and 70.7 mV
rms
(200 mV
p-p
).
The power splitter divides the 200 mV
p-p
so that 100 mV
p-p
is applied
to each of the oscilloscope inputs.
c
Press
Autoscale
.
d
Press
Source
. Set the trigger source to
Ext
.
e
Check for stable triggering, and adjust the trigger level if necessary.
f
Record results in the Performance Test Record as Pass or Fail.
If test fails, refer to "Troubleshooting the Oscilloscope."
g
Change the output of the signal generator to 25 MHz and 35.4 mV
rms
(100 mV
p-p
).
h
Check for stable triggering, and adjust the trigger level if necessary.
i
Record results in the Performance Test Record as Pass or Fail.
If test fails, refer to "Troubleshooting the Oscilloscope."
Testing, Adjusting, and Troubleshooting
To verify trigger sensitivity
7-26
Summary of Contents for 54645A
Page 5: ...iv ...
Page 12: ...1 Getting Started ...
Page 25: ...1 14 ...
Page 26: ...2 Front Panel Overview ...
Page 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Page 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Page 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Page 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Page 189: ...6 14 ...
Page 190: ...7 Testing Adjusting and Troubleshooting ...
Page 242: ...8 Replacement Parts ...
Page 259: ...8 18 ...
Page 260: ...9 Performance Characteristics ...
Page 271: ...9 12 ...
Page 272: ...10 Messages ...