To use the XY display mode
The XY display mode converts the oscilloscope from a volts-versus-time
display to a volts-versus-volts display. You can use various transducers so
the display could show strain versus displacement, flow versus pressure, volts
versus current, or voltage versus frequency. This exercise shows a common
use of the XY display mode by measuring the phase shift between two signals
of the same frequency with the Lissajous method.
1
Connect a signal to channel 1, and a signal of the same frequency but
out of phase to channel 2.
2
Press
Autoscale
, press
Main/Delayed
, then press the
XY
softkey.
3
Center the signal on the display with the Vertical Position
(HP 54645A) or Analog Position (HP 54645D) knobs, and use the
Volts/Div knobs and the vertical
Vernier
softkeys to expand the signal
for convenient viewing.
sin
θ
=
A
B
or
C
D
Example of Centering a Signal on the Display
Figure 5-8
Making Measurements with HP 54645A/D Oscilloscopes
To use the XY display mode
5-14
Summary of Contents for 54645A
Page 5: ...iv ...
Page 12: ...1 Getting Started ...
Page 25: ...1 14 ...
Page 26: ...2 Front Panel Overview ...
Page 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Page 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Page 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Page 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Page 189: ...6 14 ...
Page 190: ...7 Testing Adjusting and Troubleshooting ...
Page 242: ...8 Replacement Parts ...
Page 259: ...8 18 ...
Page 260: ...9 Performance Characteristics ...
Page 271: ...9 12 ...
Page 272: ...10 Messages ...