1
Check the rear panel
CALIBRATOR
output level.
If you are not sure how to check the
CALIBRATOR
, see "To check the output
of the
CALIBRATOR."
2
Load the default calibration factors.
a
Set the rear-panel
CALIBRATION
switch to the
UNPROTECTED
position.
b
Press
Print/Utility
. Then press
Service Menu
and
Self Cal Menu
.
c
Press the
Load Defaults
softkey.
Vertical
self cal
3
After the message "Default scope calibration factors have been
loaded" is displayed on the lower side of the display, press
Vertical
.
4
Follow the instructions on the display, then press
Continue
.
5
After the message "Press Continue to return to calibration menu"
appears on the display, press the
Continue
softkey
Delay self cal
for HP 54645A
Only
6
Connect a pulse generator set to 100 kHz and 1 V
p-p
, with a rise time
of less than 10 ns, to channels 1 and 2. Place the feedthrough at the
BNC tee. Make sure you use the HP 10502A cables to ensure equal
cable lengths.
7
Press the
Delay
softkey, then follow the instructions on the display.
8
Set the rear-panel
CALIBRATION
switch to
PROTECTED
.
Delay self cal
for HP 54645D
Only
9
Connect channels A1 and A2 to the front-panel probe adjust signal
(5 V
p-p
, 1.2 kHz) using two 10:1 probes (HP 10074A). Connect the
ground leads on the 10:1 probes to the front-panel ground lug using
the banana ground plug (HP 1252-7310).
10
Connect digital channel D0 to the front-panel probe adjust signal
(5 V
p-p
, 1.2 kHz) using the 16-channel cable (HP 54620-61801).
Connect the ground lead of this cable to the front-panel ground lug.
11
Press the Delay softkey, then follow the instructions on the display.
12
Set the rear-panel
CALIBRATION
switch to
PROTECTED
.
Testing, Adjusting, and Troubleshooting
To perform the self-calibration
7-34
Summary of Contents for 54645A
Page 5: ...iv ...
Page 12: ...1 Getting Started ...
Page 25: ...1 14 ...
Page 26: ...2 Front Panel Overview ...
Page 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Page 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Page 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Page 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Page 189: ...6 14 ...
Page 190: ...7 Testing Adjusting and Troubleshooting ...
Page 242: ...8 Replacement Parts ...
Page 259: ...8 18 ...
Page 260: ...9 Performance Characteristics ...
Page 271: ...9 12 ...
Page 272: ...10 Messages ...