![HP 54645A User'S And Service Manual Download Page 123](http://html.mh-extra.com/html/hp/54645a/54645a_users-and-service-manual_166177123.webp)
High-Frequency Probe Equivalent Circuit
The impedance plots for the two models are shown in these figures. By
comparing the two plots, you can see that both the series tip resistor and the
cable’s characteristic impedance extend the input impedance significantly.
The stray tip capacitance, which is generally small (1 pF), sets the final break
point on the impedance chart.
Impedance versus Frequency for Both Probe Circuit Models
The HP 54645D logic probes are represented by the high-frequency circuit
model shown above. They are designed to provide as much series tip
resistance as possible. Stray tip capacitance to ground is minimized by the
proper mechanical design of the probe tip assembly. This provides the
maximum input impedance at high frequencies.
Ω
Ω
Ω
Figure 4-14
Ω
Ω
Ω
Ω
Ω
Ω
Figure 4-15
Run/Stop/Single/Autostore/Erase Operation
Using Digital Channels to Probe Circuits
4-40
Summary of Contents for 54645A
Page 5: ...iv ...
Page 12: ...1 Getting Started ...
Page 25: ...1 14 ...
Page 26: ...2 Front Panel Overview ...
Page 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Page 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Page 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Page 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Page 189: ...6 14 ...
Page 190: ...7 Testing Adjusting and Troubleshooting ...
Page 242: ...8 Replacement Parts ...
Page 259: ...8 18 ...
Page 260: ...9 Performance Characteristics ...
Page 271: ...9 12 ...
Page 272: ...10 Messages ...