7 MAINTENANCE
131
7.4
Parts Maintenance and Replacement
7.4.1
Replacing the mesh filter
1.
Make sure that the sample water feed valve is closed, and the sample water is
being discharged.
Reference
Stopping operation" (page 128)
2.
Remove the filter cap.
You do not have to remove the tubes.
3.
Remove the filter stopper.
Hold a cloth under the filter holder, as water will run from the filter holder, and the filter
support may also fall.
4.
Pull out the mesh filter inside the filter holder and filter support.
5.
Wipe any dirt off of the filter stopper with a lint-free cloth, or rinse with tap water.
6.
Replace with a new mesh filter and reassemble the unit.
The direction in which you install the mesh filter and filter support is not important.
Filter holder
Filter cap
Filter support
Mesh filter
Filter stopper
Sample water feed valve
must be closed.