4 CALIBRATION
54
4.7.2
Individual zero calibration <conductivity>
This is an open zero calibration. Perform calibration only when an abnormality occurs.
Calibration procedure
1.
Close the sample water feed valve.
2.
Start the water discharge action operation.
Reference
3.
Turn off the power.
4.
Open the inner door of the device.
Reference
See "2.1 Opening the Inner Door of the Device" (page 7).
MEASUREMENT screen CALIBRATION screen COND. CAL. screen
See "4.3 Displaying the Calibration Screen" (page 23).
[CAL.] [COND.]
Sample water feed
valve must be closed.