5 FUNCTION
79
5.3.1
Setting ON/OFF of ZERO CAL.
1.
Press the ZERO CAL. button on the CALIBRATION SETTING1/2 screen.
The ZERO CAL. SETTING screen is displayed.
2.
Select [ON] or [OFF], and press [SET].
The screen returns to the CALIBRATION SETTING1/2 screen.
5.3.2
Setting the calibration interval
1.
Press the CAL. INTERVAL button on the CALIBRATION SETTING1/2 screen.
The CAL. INTERVAL SETTING screen is displayed.
2.
Enter the zero-calibration interval value using the numerical key pad and press
[SET].
Input range: 1 to 9999 (hours)
Tip
[CLEAR]: Clears the entire value.
[BACK]: Deletes the right-most number of the value.
The screen returns to the CALIBRATION SETTING screen.
Displaying CALIBRATION SETTING screen" (page 78).
MEAS. screen
MAINTENANCE
screen
SETTING1/2
screen
[SETTING]
[MAINTE.]
[CALIBRATION]
[CAL.
SETTING 1/2]
screen
[ON]
/[OFF]
Displaying CALIBRATION SETTING screen" (page 78).
MEAS. screen
MAINTENANCE
screen
SETTING1/2
screen
[SETTING]
[MAINTE.]
[CALIBRATION]
[CAL.
SETTING 1/2]
screen
CAL. INTERVAL
[xxxxhours]