4 CALIBRATION
30
4.5.1
Common zero calibration and span calibration <color>
Preparation
Rubber plug (included)
Calibration tank…1
Span calibration solution…1 L (Pour the solution into the calibration tank)
Used fluid tank (Install it so that used fluid goes into the tank)
Prepare span calibration solution according to the measurement range (0 CU to 10 CU or 0
CU to 20 CU).
Tip
Select 0 CU to 10 CU or 0 CU to 20 CU for the measuring range. See "5.9.2 Setting the range for
turbidity and color" (page 94).
Calibration procedure
Perform calibration in the order: common zero calibration
→
span calibration.
MEASUREMENT screen CALIBRATION screen COL. CAL. screen
See "4.3 Displaying the Calibration Screen" (page 23).
[CAL.] [COL.]