3 OPERATION
14
3.2.2
When stopping the device for a week or more
1.
Close the sample water feed valve.
2.
Start the water discharge operation action.
Reference
3.
Turn the power OFF.
4.
Open the inner door of the device.
Reference
"2.1 Opening the Inner Door of the Device" (page 7)
5.
Make sure that a sample water is flowing out from the turbidity/color cell and the
residual-chlorine/conductivity cell (conductivity is optional).
Tip
If water does not discharge properly, run sample water again and repeat the procedure from step 1.
6.
Remove the residual-chlorine sensor and the sample water and beads inside the
cell using the provided syringe, and then install the residual-chlorine sensor again.
Reference
"7.4.3 Maintenance of the residual-chlorine sensor" (page 133)
7.
Remove the pH and conductivity sensors (both optional) and cover the cell with a
cloth to keep it free of dust.
Reference
"7.4.4 Replacing and storing the conductivity sensor (optional unit)" (page 135)
"7.4.5 Replacing and storing the pH sensor (optional)" (page 136)
8.
Close the inner door and door of the device.
Residual-chlorine/conductivity cell
Turbidity/color cell