4 CALIBRATION
57
4.7.3
Adjustment calibration using sample water <conductivity>
Preparation
Conductivity meter...1
Calibration procedure
1.
Measure electric conductivity of the prepared sample water using a conductivity
meter other than the device's and calculate the value.
2.
Press the SAMPLE VALUE button ([xx.x]) on the COND. CAL. screen.
The sample water calibration value input screen is displayed.
3.
Enter the measured value of the sample water using the numerical key pad and
press [SET].
Input range: 0.0 mS/m to 50.0 mS/m
Tip
[CLEAR]: Clears the whole value.
[BACK]: Deletes the rightmost number of the value.
The screen returns to the COND. CAL. screen.
4.
Press [SAMPLE].
A confirmation screen is displayed asking whether to start calibration.
MEASUREMENT screen CALIBRATION screen COND. CAL. screen
See "4.3 Displaying the Calibration Screen" (page 23).
[CAL.] [COND.]
[xx.x]