4 CALIBRATION
48
4.6.3
Adjustment calibration using sample water <turbidity>
Preparation
Turbidity meter...1
Calibration procedure
1.
Measure turbidity of the prepared sample water using a turbidity meter other than
this device's and calculate the value.
2.
Press the SAMPLE VALUE button ([x.xx]) on the TURB. CAL. screen.
The sample water calibration value input screen is displayed.
3.
Enter the measured value of the sample water using the numerical key pad and
press [SET].
Input range: 0.00 FTU to 4.00 FTU
Tip
[CLEAR]: Clears the whole value.
[BACK]: Deletes the rightmost number of the value.
The screen returns to the TURB. CAL. screen.
4.
Press [SAMPLE].
A confirmation screen is displayed asking whether to start calibration.
MEASUREMENT screen CALIBRATION screen TURB. CAL. screen
See "4.3 Displaying the Calibration Screen" (page 23).
[CAL.] [TURB.]
[x.xx]