5 FUNCTION
76
2.
Select the items by using the buttons, and press [SET].
Returns to the I/O SETTING screen.
Setting the PRESET value
1.
Press [PRESET] in the lower-right portion of the screen.
The corresponding PRESET screen is displayed.
2.
Enter the PRESET value using the numerical key pad and press [SET].
Input range: 0.0 mA to 21.0 mA
Item
Explanation
Initial Setting
HOLD
Hold and output the last measured value.
If there is no last measured value when
turning ON the power, the last value
measured before turning OFF the power is
held.
DIRECT
Outputs the actual measured value.
PRESET
Output a set value.
The log data will contain the value that is
equivalent to the set value.
MAINTE./SEQ./ALARM: 21.0 mA
WARMING UP: 4.0 mA
[PRESET]