5 FUNCTION
81
5.3.5
Setting the calibration time
1.
Press the ZERO CAL. TIME button on the CALIBRATION SETTING 2/2 screen.
The ZERO CAL. SETTING screen is displayed.
2.
Enter the zero-calibration time value using the numerical key pad and press [SET].
Input range: 20 to 99 (min)
Tip
[CLEAR]: Clears the entire value.
[BACK]: Deletes the right-most number of the value.
The screen returns to the CALIBRATION SETTING screen.
Displaying CALIBRATION SETTING screen" (page 78).
MEAS. screen
MAINTENANCE
screen
SETTING1/2
screen
[SETTING]
[MAINTE.]
[CALIBRATION]
[CAL. SETTING 1/2] screen
[CAL. SETTING 2/2] screen
[
]
ZERO CAL. TIME
[xxmin]