4 CALIBRATION
38
4.5.3
Adjustment calibration using sample water <color>
Preparation
Color meter..1
Calibration procedure
1.
Adjust the flow rate of the sample water to 80 mL/min using the valve.
2.
Measure color of the prepared sample water using a color meter other than this
device's and calculate the value.
3.
Press the SAMPLE VALUE button ([x.x]) on the COL. CAL. screen.
The sample water calibration value input screen is displayed.
4.
Enter the measured value of the sample water using the numerical key pad and
press [SET].
Input range: 0.0 CU to 20.0 CU
Tip
[CLEAR]: Clears the whole value.
[BACK]: Deletes the rightmost number of the value.
The screen returns to the COL. CAL. screen.
5.
Press [SAMPLE].
A confirmation screen is displayed asking whether to start calibration.
MEASUREMENT screen CALIBRATION screen COL. CAL. screen
See "4.3 Displaying the Calibration Screen" (page 23).
[CAL.] [COL.]
[x.x]