4 CALIBRATION
56
9.
Enter 0.0 using the numerical key pad and press [SET].
Tip
[CLEAR]: Clears the whole value.
[BACK]: Deletes the rightmost number of the value.
The screen returns to the COND. CAL. screen.
10.
Press [ZERO] on the COND. CAL. screen.
A screen is displayed asking whether to start calibration.
11.
Press [YES] to start zero calibration.
The screen shows that the calibration is in progress.
12.
When the value becomes stable, press [FIX].
A confirmation screen is displayed.
13.
Press [YES].
The conductivity zero calibration coefficient is updated.
14.
Turn off the power.
15.
Install the conductivity sensor to the residual chlorine/conductivity cell.
16.
Close the inner door of the device.
17.
Open the sample water feed valve.
18.
Turn ON the power.