4 CALIBRATION
51
4.7.1
Span calibration <conductivity>
Preparation
Calibration tank…1
Span standard solution…1 L (Pour the solution into the calibration tank)
Used fluid tank (Install it so that used fluid goes into the tank)
Calibration procedure
1.
Place the calibration tank with span calibration solution beside the device.
2.
Connect the tank to the calibration inlet of the device.
3.
Press the SPAN VALUE button ([xx.x]) on the COND. CAL. screen.
The span calibration value input screen is displayed.
MEASUREMENT screen CALIBRATION screen COND. CAL. screen
See "4.3 Displaying the Calibration Screen" (page 23).
[CAL.] [COND.]
Calibration
solution tank
Connect
Sample water feed valve
must be closed.
[x.xx]