9.3 Filter Maintenance
The CM4 utilizes filters to protect the unit from dust particles and potentially damaging gases. Table 9-2 provides maintenance information
about filters. Figure 9-6 shows the relative positions of the various filters. See
for information on external sample line filters.
Name of Filter
Location
Part Number
Replacement
Frequency
Notes
Acid Scrubber Filter
Under Chemcassette
®
carrier
See
710235
6 Months
Replace the filter if the filter media (all or any) part turns
blue or sample flowrate cannot be achieved
Particulate Filters
Inside analyzer under Chemcassette
®
carrier
See
780248
3 to 6 Months
Charcoal Freon Filters
(CM4-P only)
Inside analyzer under Chemcassette
®
carrier
See Figure 9-6
1874-0139
1 Month
External Sample Line
Particulate Filter
End of sample line
(if installed)
Gas specific - See
and
Teflon Filter Housing
Assembly
End of sample line
(if installed)
1830-0055
(filter ass’y)
1 Month
Use filter membrane P/N 0235-1072 (100 per pack) or
replacement in filter housing
Table 9-2: Filter Replacement Schedule