E.7 Serial Device Applications
The serial interface is designed for unidirectional communication from
a CM4 monitor to a single printer or other remote device. It is possible,
therefore, that the application for this interface requires connection to a
remote device other than the printer options described in this appendix.
CAUTION
Cabling used between the CM4 monitor and the external device must
be shielded and properly grounded to maintain reliable operation and
control RF emissions and susceptibility
For these applications, note the pin-out configuration in each of the diagrams.
Multiple CM4 monitors and a single computer
CM4
Monitor
COM2
CM4
Monitor
COM2
CM4
Monitor
COM2
CM4
Monitor
COM2
Female DB-9
on CM4
TX 2
RX 3
Signal Gnd 5
Cable Pinout
Computer with multiple
Communication Ports
Interface
RS-232 P/N 874022-R Rear Panel Mount
Jumpers
See
for DCE settings
Computer Port Customer-supplied RS-232C; check user device for compatibility before
ordering.
Cable
User-supplied. Refer to diagram for pin-out configuration.
Also, refer to the EMC Considerations section on page
.
Protocol
Customer-written per information in Appendix E.
Note
Maximum range: 50 Ft.
Figure E-1: Serial Interface, Multiple CM4s, Single Computer, Unidirectional
Communications
Multiple CM4 monitors and a single computer
Bi-Directional Communication up to 4000 ft distance
CM4
Monitor
RS-422
CM4
Monitor
RS-422
CM4
Monitor
RS-422
CM4
Monitor
RS-422
Computer with multiple
Communication Ports
Female DB-25
Male DB-9
25 TXD+
19 TXD-
15 RXD-
17 RXD+
RXD+ 2
RXD- 3
TXD- 4
TXD+ 8
Cable Pinout
Ground 7
1 Ground
For ground, use another conductor in the cable;
do not use the shield or drain wire
Interface
RS-232 P/N 874024-R Rear Panel Mount
Computer Port Customer-supplied; check user device for compatibility before ordering.
Cable
User-supplied. Refer to diagram for pin-out configuration.
Also, refer to the EMC Considerations section on page
Protocol
Customer-written per information in Appendix E.
Note
Maximum range: 4000 Ft.
Figure E-2: Bidirectional Communication, Single Computer, Multiple CM4s
up to 4000 ft distance.