CM4 Four-Point Continuous Monitor
CM4 Technical Manual
5-2
5.1 Introduction
Use the Calibration Mode to:
• Perform tests
• Make adjustments to Chemcassette Analyzer calibration
The Calibration Mode offers four functions:
1 CALIBRATION MODE
1.1
OPTICS VERIFY
1.2
MANUAL K-FACTOR
1.3
CAL CURRENT LOOP
1.4
TUNE CURRENT LOOP
The following is an overview of each of the four functions. The following
sections describe each of these options in detail:
• 1.1 OPTICS VERIFY verifies proper operation of the
Chemcassette optics system (see Section 5.2)
• 1.2 MANUAL K-FACTOR permits entry of a new K-factor
for each particular point (see
)
• 1.3 CALIBRATE CURRENT LOOP controls the optional
external analog output for testing and installation (see
• 1.4 TUNE CURRENT LOOP controls the adjustment of
the CM4 monitor’s output level (see
)
5.2 Function 1.1 – Optics Verify
CM4 OPTICS TEST CARD
INSTRUCTIONS
FOR OPTICS VERIFICATION PROCEDURE,
ENTER CALIBRATION MODE 1.1, OPTICS VERIFY
SEE MANUAL FOR DETAILS.
STAIN 1
INSER
T
Figure 5-2: Verification Card
Purpose
Use the Optics Verify function and the Optics Verification Card to verify
proper operation of the Chemcassette Analyzer’s optics system. The
CM4 compares the nominal stain values stored in the instrument’s
memory to the values the Optics Verification Card produces.
Instructions
1. Press PROGRAM to exit Monitor Mode.
2. From the 1–CALIBRATION MODE, choose the subfunction
1–OPTICS VERIFY.
1 – CALIBRATION MODE
1 – OPTICS VERIFY
INSERT STAIN 1 PRESS