11.3 Error Displays and Remedies
181
11
The following messages are displayed when the instrument detects an error or abnormal measurement set-
ting. If repair is necessary, contact your dealer or Hioki representative.
11.3 Error Displays and Remedies
Display
Description
Remedy
+OvrRng/-OvrRng
Measurement out of range (p. 38) Select the appropriate range.
C.E. Hi
Hi wiring contact error (p. 38)
Check for cable breakage and worn
out probes.
C.E. Lo
Lo wiring contact error (p. 38)
Check for cable breakage and worn
out probes.
C.E. Volt
Voltage Level Monitor error
(p. 38)
Check for worn out probes.
ERR:001
LOW limit is higher than UPP
limit
Upper limit is below lower limit
Check comparator settings (p. 34).
ERR:002
[when LP resistance mea-
surement is disabled]
Exceeding range (From 0
to
120M
)
[when LP resistance mea-
surement is enabled]
Exceeding range (From 0
to
1200
)
Comparator input is out of range.
(out-of-range reference value,
such as 1000 M
)
Check comparator settings (p. 34).
ERR:003
Setting monitor error.
(COMP)
Comparator settings are different
from the other RM3542.
Make comparator settings the same
(p. 53).
ERR:004
Setting monitor error.
(SPEED)
SPEED setting is different from
the other RM3542.
Make SPEED settings the same
(p. 53).
ERR:011
Zero adjustment error. Offset
value exceeds 10
.
Out of zero-adjust range
Check the zero-adjustment proce-
dure (p. 32).
ERR:021
Probe short error
Short-circuited probe
Check for correct connections
(p. 51).
ERR:031
Command error
Command Error
Check for incorrect commands
(p. 115).
ERR:032
Execution error
Execution Error
Check the last command and instru-
ment state (p. 115).
ERR:033
RS-232C communication
error
I/F Communications Error
Check communication settings and
wiring.
ERR:034
Setting monitor communica-
tion error
Settings monitor communications
error
Check settings monitor settings and
wiring (p. 53).
ERR:041
Line frequency detection
error
Error detecting the line frequency
Manually set the frequency to match
the line (p. 63).
ERR:042
Clock error
Reset?(09-01-01 00:00:00)
Press F1 Key.
The clock is not set, so pressing
F1 [OK] displays the initialized
time 09-01-01 00:00:00.
The back-up battery needs to be re-
placed. Contact your nearest Hioki
representative.
ERR:101
Hardware error
(Main CPU ROM)
Hardware failure.
Repair is required.
ERR:102
Hardware error
(Main CPU RAM)
Hardware failure.
Repair is required.
ERR:103
Hardware error (SRAM)
Hardware failure.
Repair is required.
ERR:104
Hardware error
(Adjustment data)
Hardware failure.
Repair is required.
ERR:105
Hardware error
(Backup data)
Reset? Press F1 Key.
Hardware failure. (back-up data is
corrupted)
Repair is required.
Press F1 [OK] to resets the instru-
ment.
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......