9.7 Message Reference
138
(7) Triggering
Relationship Between Trigger Source and Continuous Measurement Operation
Operation depends on the continuous measurement setting (
:INITIATE:CONTINUOUS
,
the trigger source setting (
:TRIGGER:SOURCE
,
), as follows.
See:
"9.8 Data Exporting Methods" (p. 148)
Measurement Flow
Continuous Measurement Command-Specific Settings
:INITIATE:CONTINUOUS ON
:INITIATE:CONTINUOUS OFF
Trigger
Source
:TRIGGER:SOURCE IMM
Free-Run state.
Measurement continues automatically.
Trigger by
:INITIATE
(or
:READ?
) com-
mand.
:TRIGGER:SOURCE EXT
Trigger by TRIG signal,
F4
[MANU] key or
TRG
command. After measurement, enters
the trigger wait state.
Issue
:INITIATE
(or
:READ?
) command
to wait for trigger. Trigger by TRIG signal.
:INITIATE:CONTINUOUS OFF
Can only be set by Remote command.
If this has been set to OFF when operation is returned to the Local state or power is turned off, the following state occurs
when power is turned back on.
:INITIATE:CONTINUOUS ON
("To cancel the Remote state (enter the local state)" (p. 104)
Exporting measured values: "9.8 Data Exporting Methods" (p. 148)
Measurement
Calculation
Measured Value Output
Trigger Delay
Idle State
Measurement
Calculation
Measured Value Output
Trigger Delay
:INITIATE:IMMEDIATE
or
:READ?
Trigger Wait State
Measurement
Calculation
Measured Value Output
Trigger Delay
TRIG signal, F4 [MANU]
key,
TRG
Trigger Wait State
Measurement
Calculation
Measured Value Output
Trigger Delay
Idle State
TRIG signal
:INITIATE:IMMEDIATE
or
:READ?
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......