4.6 Detecting Measurement Voltage Faults (Voltage Level Monitor Function)
49
4
When a measurement voltage fault occurs due to probe chattering, the
C.E. Volt
error message
appears on the measurement screen and an
ERR
signal is output.
The
C.E. Volt
error may also appear when external noise is strong.
Check the following if errors occur frequently:
• Check for probe deterioration.
• Provide additional noise suppression. "Appendix 3 Unstable Measurement Values" (p. A3)
• Set the voltage level monitor to Loose, or OFF (disable).
4.6
Detecting Measurement Voltage Faults
(Voltage Level Monitor Function)
1
Open the Basic Settings screen.
2
Open the Measurement Settings Screen.
3
Enable or disable the function.
4
(When enabled (ON selected))
Select the voltage level monitor threshold.
5
Return to the Measurement screen.
The Basic Settings screen appears.
The Measurement Settings Screen
appears.
[MEAS SETTINGS]
1
Selection
2
Selection
1
2
Disables the function (go to step 5)
Enables the function (default)
The setting is specific to the selected range (p. 41)
Selection
1
2
Loose *
Normal*
Severe
*. Default setting: Loose is
the default for the 100 M
range, and Normal for all
other ranges.
The confirmation screen appears.
Return to the setting screen.
Save setting and return to
previous screen.
Discard setting and return to
previous screen.
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......