9.8 Data Exporting Methods
148
Basic Data Exporting Methods
Flexible data exporting is available depending on the application.
9.8
Data Exporting Methods
Free-Run Data Exporting
Initial Setup
:INITiate:CONTinuous ON
(continuous measurement enable)
:TRIGger:SOURce IMMediate
(internal triggering)
Exporting
:FETCh?
Imports the most recent measurement.
Exporting by Host Triggering
Initial Setup
:INITiate:CONTinuous OFF
(continuous measurement disable)
:TRIGger:SOURce IMMediate
(internal triggering)
Exporting
:READ?
A trigger occurs, and a measurement is taken and the result is transferred.
Exporting by Pressing the F4 [MANU] Key
*1
or Applying TRIG Signal
Initial Setup
:INITiate:CONTinuous OFF
(disable continuous measurement)
:TRIGger:SOURce EXT
(external triggering)
Exporting
:READ?
When triggered by the
F4
[MANU] Key
*1
or TRIG signal, a measurement is taken and the
result is transferred.
*1. The
F4
[MANU] key indicator is displayed when [TRG: EXT] is selected on the Basic Setting screen.
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......