9.7 Message Reference
146
Set and Query Measurement Current Mode
Syntax
Command
:RESistance:CURRent
<Range>
,
<PULSe/ CONTinuous>
Query
Response
:RESistance:CURRent?
<Range>
<Range>,<PULSE/ CONTINUOUS>
Command
:LPResistance:CURRent
<Range>
,
<PULSe/ CONTinuous>
Query
Response
:LPResistance:CURRent?
<Range>
<Range>,<PULSE/ CONTINUOUS>
<Range>
=
(:RESistance) RNG100MIL/ RNG1000MIL/ RNG10/ RNG100/ RNG1000/ RNG10K/
RNG100K/ RNG1000K/ RNG10MEG/ RNG100MEG
(:LPResistance) RNG1000MIL/ RNG10/ RNG100/ RNG1000
<PULSe>
= Measurement current is applied as a pulse only during measurement.
<CONTinuous>
= Measurement current is applied continuously.
Description
Selects the method of current application. Select CONTinuous when measuring objects that
require time for stabilization after applying measurement current.
Note
When using the Contact Improver function (set to HOLD or PULSE), measurement current is only
applied as a pulse, even if CONTinuous is selected here. Therefore, to measure with continuous
current, the Contact Improver function must be disabled (set to OFF). (p. 145)
Set and Query Contact Check Operation
Syntax
Command
:RESistance:CONTactcheck
<Range>
,
<1/ 0/ ON/ OFF>
Query
Response
:RESistance:CONTactcheck?
<Range>
<ON/ OFF>
Command
:LPResistance:CONTactcheck
<Range>
,
<1/ 0/ ON/ OFF>
Query
Response
:LPResistance:CONTactcheck?
<Range>
<ON/ OFF>
<Range>
=
(:RESistance) RNG100MIL/ RNG1000MIL/ RNG10/ RNG100/ RNG1000/ RNG10K/
RNG100K/ RNG1000K/ RNG10MEG/ RNG100MEG
(:LPResistance) RNG1000MIL/ RNG10/ RNG100/ RNG1000
Description
Enables/disables the Contact Check function.
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......