9.5 Communication Methods
114
Register Reading and Writing
GP-IB Commands
The following commands can be used for performing interface functions.
Register
Read
Write
Status Byte Register
STB?
-
Service Request Enable Register
SRE?
SRE
Standard Event Status Register
ESR?
-
Standard Event Status Enable Register
ESE?
ESE
Event Status Register 0
:ESR0?
-
Event Status Enable Register 0
:ESE0?
:ESE0
Event Status Register 1
:ESR1?
-
Event Status Enable Register 1
:ESE1?
:ESE1
Command
Description
GTL
Go To Local
Cancels the Remote state and enters the Local state.
LLO
Local Lock Out
Disables all keys, including the Local (F1) key.
DCL
Device CLear
Clears the input buffer and the output queue.
SDC
Selected
Device Clear
Clears the input buffer and the output queue.
GET
Group Execute
Trigger
When an external trigger occurs, processes one sample.
Initialization Items
Initialization Method
Item
At
Power-on
Key reset
RST
Command
Device
Clear
(GP-IB only)
CLS
Command
Factory
Default
GP-IB Address
1
1
RS-232C setting (fixed)
9600
9600
Device-specific functions
(Range, etc.)
Output Queue
Input buffer
Status Byte Register
*1
*2
Event registers
*3
Enable register
Current path
Headers on/off
OFF
OFF
OFF
OFF
Response message terminator
(GP-IB)
LF+EOI
LF+EOI
LF+EOI
Response message separator
*1. Only the MAV bit (bit 4) is cleared.
*2. All bits except the MAV bit are cleared.
*3. Except the PON bit (bit 7).
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......