3.7 Judging Measured Values (Comparator Function)
37
3
Setting example: Upper threshold 150 m
, lower threshold 50 m
To abort the setting process, press
. Settings are abandoned and the display returns to the
previous screen.
Decide According to Upper/Lower Thresholds (ABS Mode)
1
Open the absolute value threshold setting screen.
2
Set the positive tolerance.
Pressing an inoperative key during setting sounds a low-pitch beep (when the key beeper is enabled).
No error message is displayed.
Press the units key to accept the setting and move the cursor to the lower threshold value.
3
Set the negative tolerance in the same way.
Press the units key to accept the setting and move the cursor to the upper threshold.
4
Accept the settings and return to the Measurement screen.
Upper threshold
Lower threshold
1_
15_
150_
150.0m
(Example: 150 m
)
Selection
To Reset Numerical Values
Deletes entered digits.
This key is enabled only when enter-
ing numerical values.
To change the value after selecting the
units, use the cursor
keys to select
the item to change, then enter the new val-
ue with the tenkeys.
1
2
1
2
Selection
(Example: 50 m
)
5_
50_
50.0m
•
Internal calculations are performed on floating-point values, and decisions round up any fraction of the
least-significant digit.
•
Displayed values of the reference and tolerances are rounded according to the selected range. Internal
calculations use unrounded data, so decisions are based on the entered (setting) values.
•
An error message appears if you press
ENTER
with the positive tolerance set lower than the negative
tolerance.
See:
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......