4.7 Applying Current Only When Measuring (Current Mode Setting)
50
When the Contact Improver function is set to Pulse or disabled (CONT IMP: PULSE or OFF) and
measurement current is set for pulse output, open-circuit voltage when not measuring does not
exceed 20 mV.
See:
"4.5 Improving Probe Contact (Contact Improver Function)" (p. 47)
When the Contact Improver function is enabled (CONT IMP: PULSE or ON setting), the current mode setting
is ignored even if set to continuous (CURRENT MODE: CONT setting). The Contact Improver function forces
pulse operation with measurement current applied only during measurement.
4.7
Applying Current Only When Measuring
(Current Mode Setting)
1
Open the Basic Settings screen.
2
Open the Measurement Settings Screen.
3
Select whether to apply current when not measuring.
To apply measurement current continuously (CONT setting) even when waiting for a trigger, confirm that
the Contact Improver function is disabled (CONT IMP: OFF, (p. 47)).
4
Return to the Measurement screen.
The Basic Settings screen appears.
The Measurement Settings Screen
appears.
[MEAS SETTINGS]
1
Selection
2
Selection
1
2
Measurement current is applied while
awaiting trigger.
Measurement current is applied only
while measuring (default).
The setting is specific to the selected range (p. 41)
The confirmation screen appears.
Return to the setting screen.
Save setting and return to
previous screen.
Discard setting and return to
previous screen.
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......